|
Niels
Fogh-Andersen*, Paul D'Orazio, Katsuhiko Kuwa,
Wolf R. K�lpmann, Gerhard Mager, and Lasse Larsson
*Send comments on this document to:
Niels Fogh-Andersen, MD
Department of Clinical Biochemistry,
Herlev Hospital
DK-2730 Herlev, Denmark
email: nfa@post6.tele.dk
**For purposes of this document, direct reading biosensors
are defined
Download as a
PDF here
Different devices for the measurement of glucose detect and
report fundamentally different analytical quantities. In man,
glucose distributes like water between erythrocytes and plasma,
carried by the erythrocyte glucose transporter. Therefore, molality
of glucose (amount per unit water mass) is equal in plasma and
erythrocyte fluid. Different water concentrations in calibrator,
plasma, and erythrocyte fluid can explain some differences
dependent on sample type, methods requiring sample dilution, and
"direct reading" biosensors** detecting molality. The original
intention of the IFCC Document was to recommend reporting of plasma
equivalent glucose concentrations for direct-reading biosensors in
blood gas/electrolyte/metabolite analyzers. However, an isolated
recommendation will not lead to globally commutable results, which
require a consensus on reporting results from all analyzers. The
IFCC-SD Working Group on Selective Electrodes recommends
harmonizing to the concentration of glucose in plasma (with the
unit mmol/L), irrespective of sample type or measurement technology
used.
Glucose permeates the erythrocyte membrane quickly, by passive
transport (facilitated by the erythrocyte glucose transporter,
which catalyzes the uniport movement of D-glucose down its
concentration gradient). Therefore, glucose distributes between
erythrocytes and plasma like water. The activity (or molality) of
glucose inside erythrocytes equals that in plasma, providing
equivalent results for blood and plasma when measured with a direct
reading glucose biosensor. Activity (of symbol a, without unit) is
related to the chemical potential (� = �0 + RTln a of unit kJ/mol)
used in calculations of free energy changes, reaction affinity etc.
The activity of glucose is assumed equal to molality, or amount per
unit water mass, m of unit mmol/kg H2O. Activity is the
physiologically relevant quantity, determining enzymatic reaction
rate, direction of chemical processes, transport, binding to
receptors etc. The activity (or molality) of glucose in blood is
physiologically relevant, but not recommended as a quantity for
clinical use in this document. A new quantity like activity or
molality of glucose in plasma would only increase the present risk
of clinical misinterpretation and add to the confusion regarding
sample type and measurement technology.
Various types of instruments now detect and report fundamentally
different glucose quantities. Inexpensive instruments with
direct-reading biosensors are widely available for self-monitoring
or point of care glucose testing (1-3). For the foreseeable future,
the clinical chemistry laboratory is expected to perform glucose
determinations by direct reading sensors concurrently with other
routine instruments. Unlike direct reading glucose biosensors that
detect molality, sensors that require diluted samples produce
results that depend on water concentration of the sample. On a
concentration basis (amount of glucose per liter of sample),
glucose in plasma is higher than glucose within erythrocytes,
because the water concentration is higher in plasma than in
erythrocytes. Therefore, biosensors relying on sample dilution will
produce higher results for plasma than the corresponding blood, by
approximately 11% for blood of normal hematocrit. Furthermore, the
clinical staff in general does not know whether the laboratory
results are for blood or plasma glucose (4).
The World Health Organization (WHO) (5, 6) and American Diabetes
Association (ADA) (7) define diabetes mellitus by more than one
fasting plasma glucose concentration > 7.0 mmol/L. As an
alternative, a casual plasma glucose concentration > 11.1 mmol/L
in the presence of symptoms or a 2-h post oral glucose tolerance
test result > 11.1 mmol/L suffice to make a definite diagnosis
of diabetes mellitus. The new category of "impaired fasting plasma
glucose concentration" has a narrower interval of 6.1-6.9 mmol/L
than the previous fasting interval of 5.6-7.7 mmol/L between normal
and diabetic classifications. With the present use of multiple
methods providing different results, there is a serious risk for
clinical misinterpretation. The WHO and ADA categorize patients
based on their plasma glucose concentration. The ADA further
recommends no more than 5 % analytical error for future glucose
monitors, with a maximum of 15% total error and 10 % imprecision
(8, 9). The systematic 11 % difference between normal blood and
plasma glucose concentration alone exceeds the maximum analytical
error. We recommend always reporting the concentration of glucose
in plasma to avoid ambiguity. The choice of plasma instead of whole
blood is somewhat arbitrary. However, leading practitioners in the
field of diabetes management prefer plasma glucose concentration as
the quantity of choice (from personal communication with KGMM
Alberti). When whole blood glucose concentration is measured, a
constant factor of 1.11 will convert whole blood to plasma glucose
concentration. The factor 1.11 is the ratio of water and therefore,
glucose concentrations in normal plasma and whole blood. We
recommend always using a constant factor of 1.11. An individual
conversion based on hematocrit may introduce additional imprecision
(10), besides being less convenient and requiring additional
information. The converted plasma glucose concentrations will have
the same dependence on hematocrit as the presently reported whole
blood glucose concentrations.
Consider, e.g., a blood specimen with a normal hematocrit (Hct) of
0.43. The water concentration of erythrocytes is ~ 0.71 kg/L. The
water concentration of plasma is ~ 0.93 kg/L. The water
concentration (kg H2O/L) of the blood specimen must be
intermediate, (0.43)*(0.71)+(1-0.43)*(0.93) = 0.84. The ratio of
water (and therefore, glucose) concentration between plasma and
whole blood is 0.93/0.84, or 1.11. The ratio depends on hematocrit.
A decreased Hct causes an increased glucose concentration in whole
blood and vice versa. When hematocrit is known to be abnormal,
whole blood glucose concentration may be "hematocrit adjusted" to a
normal hematocrit of 0.43 by multiplication with
0.84/(0.93-0.22*Hct). Unfortunately, some methods may have
additional erythrocyte or hemoglobin interference.
Direct-reading glucose biosensors detecting molality of glucose in
whole blood are available on combined blood
gas/electrolyte/metabolite analyzers from all the major
manufacturers of these systems. Most of these systems presently
correlate to the plasma equivalent concentration of glucose.
However, one manufacturer calibrates with aqueous calibrators
without considering the different concentrations of water in sample
and calibrator, providing 'relative molality' of glucose in the
sample. The predicted ratio of results for unmodified
direct/diluted methods is 0.99/0.84 = 1.18 for whole blood and
0.99/0.93 = 1.06 for plasma, in harmony with results from the
literature (10, 11). Most devices for point of care analysis of
blood glucose also use direct reading biosensors, and some
calibrate to the plasma equivalent concentration of glucose (1).
Continued use of the variety of systems presently available for
measurement of glucose without conversion to plasma results may
cause confusion with conventionally measured and reported glucose
concentrations, for example from the central laboratory. All
reference intervals and clinical decision levels must accordingly
reflect plasma results. Figure 1
References
1. Chance JF, Li DJ, Jones KA, Dyer KL, Nichols JH. Technical
evaluation of five glucose meters with data management
capabilities. Am J Clin Pathol 1999; 111: 547-56.
2. Johnson RN, Baker JR. Accuracy of devices used for
self-monitoring of blood glucose. Ann Clin Biochem 1998; 35:
68-74.
3. Kost GJ et al. Multicenter study of oxygen-insensitive
handheld glucose point-of-care testing in critical
care/hospital/ambulatory patients in the United States and Canada.
Crit Care Med 1998; 26: 581-9.
4. Burrin JM, Alberti KGMM. What is blood glucose: can it be
measured? Diabetic Med 1990; 7: 199-206
5. Alberti KGMM, Zimmet PZ. Definition, diagnosis and
classification of diabetes mellitus and its complications. Part I:
diagnosis and classification of diabetes mellitus. Provisional
report of a WHO consultation. Diabetic Medicine 1998; 15:
539-53.
6. Diabetes mellitus. Report of a WHO Study Group, World Health
Organization Expert Committee. Tech Report Ser 727. Geneva,
Switzerland: WHO, 1985.
7. Report of the expert committee on the diagnosis and
classification of diabetes mellitus. Diabetes Care 1997; 20:
1183-97.
8. American Diabetes Association. Consensus statement - self
monitoring of blood glucose. Diabetes Care 1990; 13 (Suppl 1):
41-6.
9. American Diabetes Association. Self-monitoring of blood
glucose. Diabetes Care 1994; 17: 81-6.
10. Fogh-Andersen N, D'Orazio P. Proposal for standardizing
direct-reading biosensors for blood glucose. Clin Chem 1998; 44:
655-9.
11. Fogh-Andersen N, Wimberley PD, Thode J, Siggaard-Andersen O.
Direct reading glucose electrodes detect the molality of glucose in
plasma and whole blood. Clin Chim Acta 1990; 189: 33-8.
|