|
Neil
Greenberg, PhD, DABCC, Manager, Regulatory Affairs, Ortho-Clinical
Diagnostics, Inc., A Johnson & Johnson Company
This article is published with permission and originally
appeared:
Neil Greenberg, "Calibrator Traceability: The Industry Impact of
the IVD Directive's New Requirements," IVD Technology 7, no. 2
(2001): 18-27
Download as a
PDF here
Introduction
Traceability to internationally recognized and accepted
standards is an important component in assuring the accuracy and
comparability of clinical laboratory measurements. Currently, the
global marketplace is presenting new demands on IVD device
manufacturers for measurement traceability. Under a mandate from
the European Union�s In Vitro Diagnostics Directive1 (IVDD), the
European Committee for Standardization�s Technical Committee 140
(CEN/TC 140 - In Vitro Diagnostic Systems), together with ISO/TC
212, is currently developing international standards on IVD
calibration traceability2,3. Full implementation of the IVD
Directive, which is obligatory by December 2003 under European law,
will require that calibration of quantitative IVD assays be
traceable to available �higher-order� reference materials or
methods. Manufacturers who implement the processes and
documentation steps defined in the calibration traceability
standards are entitled to a presumption of conformity to this
�essential� requirement of the IVD Directive. To do this, IVD
manufacturers must ensure that the systems they market have been
calibrated against available higher-order reference standards and
procedures, that repeatability and reproducibility of their
internal calibration procedures are quantified and documented, and
that accuracy is substantiated by uncertainty calculations.
The draft international standard for IVD calibration
traceability, ISO/DIS 17511, identifies the essential elements of a
calibration hierarchy necessary to support full calibration
traceability to defined units of measure under the Syst�me
International (e.g. moles, kilograms). In addition to a
complete definition of the quantity to be measured, there is a need
for primary standards, including material standards as well as
standard methods of measurement. In the introductory section
of the text, the standard also discusses a key problem that exists
for measurement systems in the field of in vitro diagnostics. This
is that, although the in vitro diagnostics field routinely performs
measurements on an estimated 400 to 600 different amounts of
substances (analytes), full calibration systems with traceability
to SI currently exist for less than 30 (perhaps 5%) of these
analytes. What will be done to fill this 95% void, and what
role will be assumed by the IVD industry?
Do IVD manufacturers
want additional international reference materials and methods to be
developed?
Despite short-term concerns, international IVD calibration
standards (reference systems) ultimately help IVD manufacturers by
providing well-defined market needs and customer requirements, a
clear and universal definition of goals, and tools for objective
assessment of product attributes. Table 1 lists a few of the global
trade-offs to be evaluated by all participating organizations
(manufacturers and distributors of IVD products, industry
associations, user groups, customer advocacy and professional
groups, government and regulatory bodies) upon considering
development of new calibration standards (reference systems) for
IVDs. It is clearly a matter of cost vs. benefit, and the decision
to invest in standardization demands careful analysis on a
case-by-case basis.
TABLE 1. Decision Trade-Offs for Investment in
New IVD Reference Systems
An Example: Economic Impact Analysis of Standard Reference
Materials for Cholesterol:
A recent economic impact study conducted by the US National
Institute of Standards and Technology (NIST) quantified a portion
of the economic benefits associated with the availability of NIST
Cholesterol Standard Reference Materials beginning in 19864. This
study determined that the economic consequences of NIST�s
Cholesterol Standards Program were experienced at several levels of
the IVD medical device supply chain from manufacturers, to network
laboratories, and to clinical laboratories that ultimately deliver
medical services to the consumer. The nature of the benefits to
industry resulting from the NIST investment have changed over more
than three decades of NIST involvement. However, because the
timeframe of this analysis was limited to 1986-1999, only part of
the NIST program�s life cycle, the magnitude of the estimated
economic impact was biased low. Nevertheless, the results indicated
that NIST reference materials played an important economic role in
support of the US national effort to monitor, measure, and control
cholesterol levels, thereby contributing to a reduced level of
heart disease. The study estimated a benefit-to-cost ratio of 4.5,
and a social rate of return of 154 %. The Net Present Value was
calculated to be more than (US) $3.6 million. The study did not
attempt to account for the impact of NIST reference materials on
reducing the incidence of incorrect laboratory measurements on
patient care, which has been estimated to be approximately $100
million per year in the US.
Impact of Measurement
Bias on Globalization of Healthcare
In the 21st century, we live in an era where patients frequently
move from town to town or even country to country as often as every
few years. Additionally, for economic reasons, there is an
increase in the rate at which patients change to different health
care plans and different clinical laboratories. As the
mobility of medical records increases along with patient mobility,
differences in laboratory test results among laboratories and
across different test methods are becoming increasingly apparent to
physicians and other health care practitioners.
Problems due to lack of consistency in standardization among
analytical methods were highlighted and publicly debated in the
United States during the 1980�s, following the US National
Institutes of Health publication of its findings on the
relationship between serum cholesterol levels and risk for
cardiovascular disease. Similarly, in the mid-1990�s, a
recommendation was published for monitoring the change in serum PSA
over 2 to 4 years, as an aid in the diagnosis of prostate cancer.
This recommendation made inter-laboratory variability in PSA
determinations highly problematic, since a large change in PSA
could result simply because of analytical differences in the
methods used by two labs. Alternatively, clinically
significant changes in true PSA levels could be masked for the same
reasons.
Klee5 has evaluated the cost-impact of certain biased laboratory
test results associated with unnecessary followup of mis-classified
patients. For a screening test such as serum cholesterol, patient
mis-classification is likely to be followed up with expensive
additional testing or even inappropriate treatment. Using actual
test results distributions rather than Gaussian models, Klee
estimated that an assay for serum cholesterol that is biased 1.0%
high results in a 3.0 % increase in the number of patients
classified as having �high� cholesterol values. Similarly, a
3 % bias causes an 8.8 % increase in the number of patients
classified as hypercholesterolemic, and a 10 % bias causes a 27.8 %
increase. Overall, the change in the percentage of patients
crossing the serum cholesterol decision threshold (5.17 mmol/L; 200
mg/dL) increases about three fold more than the percentage value of
the analytical shift due to the multiplier effect of the
distribution curves. Based on these estimates, improvements in
laboratory standardization (especially for screening tests for
clinical conditions with high prevalence) makes sound economic
sense, whenever the outcome is likely to be lower rates of patient
mis-classification.
Key factors leading
to successful improvement in inter-laboratory standardization
Why do some standardization programs meet with success, while
other programs seem to languish in some instances for years without
yielding noticeable benefits? Following up on the examples of
cholesterol and PSA discussed above, it is clear that the programs
to improve standardization of serum cholesterol measurements
yielded measureable success in terms of a very substantial
reduction in measurement variability among laboratories over a ten
year period. However for PSA, although some progress is
evident, success remains elusive6. Similarly, inter-method and
inter-laboratory standardization efforts for measurement of human
chorionic gonadotropin (HCG) in serum and urine, another clinically
important analyte which is an excellent marker for screening and
monitoring normal and abnormal pregnancy as well as certain
malignancies, have also encountered significant technical
barriers7.
In a recent review of the history of standardization efforts in
the clinical laboratory, Eckfeldt8 identified four significant
antecedents to successful laboratory standardization
programs. These include:
- Results of a widely publicized clinical research study conclude
that clinical action based on application of uniform cut-points for
a particular laboratory test leads to significant improvement in
detection and prognosis of patients with disease. This new
information leads to pressure from large clinical organizations for
improvements in test method accuracy and reliability.
- A high-level reference method and/or material exists.
- Mechanisms exist to easily and reliably disseminate the
accuracy base provided by the reference method and/or
material.
- Tools exist to reliably evaluate and publicly display
inter-method and inter-laboratory performance data. EQAS
programs are a primary source of such information, and the value of
such programs depends on the free flow of information about their
procedures and the test materials that they distribute.
The importance of the reliability of EQAS data in assessment of
the accuracy of laboratory methods is often underestimated. It is
essential that EQAS materials achieve the highest possible level of
commutability, so as to ensure their validity in representation of
performance of a given laboratory test with patient samples.
Indeed, incorrect assumptions about the validity and commutability
of EQAS materials may lead to incorrect conclusions and even
serious errors on the part of IVD kit manufacturers as well as EQAS
providers and laboratorians. In the 1980�s, before information
about commutability problems and matrix effects with manufactured
(typically lyophilized) serum controls and EQAS/PT materials was
widely publicized, there were examples where manufacturers
unwittingly adjusted the calibration of their IVD devices and
reagents to make PT/EQAS samples� results comparable to reference
method target values. In some cases, these adjustments
compromised accuracy with patient samples.
Concerns about the suitability of PT/EQAS materials have led to
uncertainty regarding the value of PT/EQAS results in understanding
the state-of-the-art for trueness. These concerns have also
been responsible for creating some tension between providers of
PT/EQAS programs and IVD manufacturers whose commercial methods are
evaluated by these programs. IVD manufacturers often respond that
inferior PT materials do not accurately simulate clinical
specimens, and mis-represent the performance of their
methods. The providers of PT/EQAS materials and programs
often argue that commercial reagent and instrument systems are
insufficiently "robust". In reality, the "problem" must be shared
by the broader professional laboratory community and the commercial
IVD industry as a whole.
Declaration of New or
Improved International Reference Materials and ReferenceMethods �
Some Manufacturers May Need to Change Calibrations for Certain
Analytes
Development of new or improved reference systems is not a
specific requirement of the IVD Directive. The Directive states
that routine methods need to be traceable to ��available reference
measurement procedures and/or available reference materials of a
higher order.� Similarly, ISO/CD 17511 allows for a wide range of
scenarios, including situations where there is no recognized higher
order calibration method or material available to trace back to.
Under these circumstances, the highest order reference point
available to a given IVD manufacturer may be a measurement
procedure or reference material that is uniquely defined,
controlled, and maintained by each manufacturer of the various
commercial assay systems for a given analyte. ISO/CD 17511 takes a
stronger position in terms of commitment to the cause of continuous
improvement in reference methods and materials. As the standard
states, �It is the aim of metrology in laboratory medicine to
improve traceability�by providing the missing reference measurement
procedures and reference materials, based on international
consensus.� Because of this implied commitment, it is expected that
many national and international standards organizations,
scientific, professional, and industry groups, will interpret the
Directive�s intent, arriving at an interpretation which says that
the Directive demands investment in upgrades to the international
reference system for the clinical lab.
When new reference methods or materials are developed and become
globally accredited, certain IVD manufacturers will inevitably have
to make changes in their internal calibration procedures in order
to adjust performance of their products to become standardized to
new reference systems9. The cost implications for these changes are
far-reaching, and will impact end-users as well. Table 2
highlights a few of these costs.
Most successful IVD device manufacturers are willing to step up
and implement calibration changes necessitated by customer needs,
especially when these changes are expected to be beneficial in
bringing about improved health care. Hopefully, the required
changes will be important ones, representing needs articulated by a
broad, global consensus of clinicians and laboratory professionals,
since many changes may be associated with disruption and costs.
Given these costs, it is especially important that the initiative
and the leadership for change be customer-focused, originating from
a clear expression of need for improvement on the part of the
end-users, not the manufacturers.
What is most important, as new reference materials or reference
method projects are initiated in the name of the Directive, is that
adequate scientific support from industry be sought when staffing
the technical working groups. Active and meaningful participation
by industry scientists will help to ensure that a reasonable
balance is achieved between commercial interests, pragmatic
realities of manufacturing materials and process limitations, and
academic clinical and metrological interests.
Does Industry Support
Initiatives for New IVD Calibration and Reference Systems?
Public interest requires good quality and safe health care
products. It is easy to demonstrate that improved standardization
contributes to furthering these goals, and all IVD manufacturing
companies share this interest. Reference materials and reference
methods for calibration are an important underlying element,
contributing to quality and safety through the assurance of
interchangeability of information across time and space.
Given these shared goals, it is inevitable that new reference
systems projects will emerge, and new reference materials and
methods will ultimately be defined, impacting the definition of the
state of the art. Industry must play a role in this process, using
its collective wisdom achieved through years of real-world
experience, to ensure technically sound and practical solutions to
the challenges encountered in development projects undertaken in
the quest for better standards.
Recommendations
What factors are necessary to get to the future state? To begin,
laboratory medicine, scientific, and professional organizations
need to provide leadership and guidance relative to what standards
are needed. In doing so, project priorities should be defined with
an appreciation that resources are limited, while taking into
account factors such as (1) the public health significance and
disease course, (2) expectations of the degree of improvement
anticipated in overall clinical effectiveness of a given test if a
new standard is developed, (3) time and cost estimated to reach a
desirable endpoint, and (4) the overall likelihood of success.
High priority should be given to establishing a defined, global,
customer-focused, and consensus-based process for setting
priorities and contracting projects. This process should be led by
a consortium of the world�s major laboratory professional
associations (e.g. IFCC, AACC, CAP, WASP and others) and should
proceed in an atmosphere of open and public dialogue, employing
decision tools that emphasize quantification and metrics, and is
inclusive of all key stakeholders (i.e., profession, government,
industry, lay public).
Sufficient project funding (grants or contracts) must be
provided, and should include government sources (e.g. EU
Commission, US Department of Commerce, etc.), professional
societies (e.g. IFCC, AACC, CAP, WASP), as well as industry groups.
Projects must be closely managed, with accountability for
deliverables and schedule. Project teams must be staffed with
appropriate clinical and scientific experts, coming equally from
the professions and industry.
Conclusions
Although the EU IVD Directive does not specifically require it,
reference materials and reference methods development projects will
be initiated in the name of the Directive�s essential requirement
for calibration traceability. New calibration standards are likely
to increase short-term costs for IVD manufacturers, but this is a
minor consideration if there is a clear need for improved standards
and an expectation of improved quality of health care as an
outcome, as expressed by a consensus of customers and professional
associations.
Professional and customer advocacy groups should take the lead
role in advocating for new calibration standards, especially in
defining where improved standards are needed. Project selection
must utilize cost-benefit analysis, taking into account public
health payback, technological limitations, and magnitude of the
investment necessary to achieve the desired outcome.
Whenever IVD calibration and reference systems standards
projects are undertaken, whether sponsored by professional,
government, or public health groups, inclusion of IVD industry
scientists and experts on the technical team is an absolute
prerequisite for project success. Creative strategies are needed to
ensure adequate project funding, and should involve a combination
of contributions from public, professional and industry
sources.
References
1. Council Directive
98/79/EC of the European Parliament and of the Council of 27
October 1998 on In Vitro Diagnostic Medical Devices, Official
Journal of the European Union L331 (December 12, 1998).
2. In Vitro Diagnostic
Medical Devices�Measurement of Quantities in Samples of Biological
Origin�Metrological Traceability of Values Assigned to Calibrators
and Control Materials, ISO/CD 17511 (Geneva: International
Organization for Standardization, February, 2000).
3. In Vitro Diagnostic
Medical Devices�Measurement of Quantities in Samples of Biological
Origin�Metrological Traceability of Values for Catalytic
Concentration of Enzymes Assigned to Calibrators and Control
Materials, ISO/CD 18153 (Geneva: International Organization for
Standardization, February, 2000).
4. May, Willie E, NIST
Measurement Methods and Standards Reference Materials for Health
Status Markers: Current Program and Future Challenges, in: NISTIR
6742, Proceedings of the Workshop on Measurement Traceability for
Clinical Laboratory Testing and In Vitro Diagnostic Test Systems,
Ellyn S. Beary, Editor, May 2001, U. S. Department Of Commerce,
National Institute of Standards and Technology.
5. Klee, George G.,
Importance of Commutable Reference Materials and Patient Test
Distributions for Assay Calibration, in: NISTIR 6742, Proceedings
of the Workshop on Measurement Traceability for Clinical Laboratory
Testing and In Vitro Diagnostic Test Systems, Ellyn S. Beary,
Editor, May 2001, U. S. Department Of Commerce, National Institute
of Standards and Technology.
6. Chan DW, Sokoll LJ.
WHO first international standard for prostate-specific antigen: the
beginning of the end for assay discrepancies? Clin Chem
2000;46:1291-2.
7. Galina
Kovalevskaya, Steven Birken, Tatsu Kakuma, John Schlatterer and
John F. O�Connor, Evaluation of Nicked Human Chorionic Gonadotropin
Content in Clinical Specimens by a Specific Immunometric Assay,
Clinical Chemistry, 1999; 45: 68-77.
8. Eckfeldt, John H.,
History of Reference Systems for Clinical Measurements, in: NISTIR
6742, Proceedings of the Workshop on Measurement Traceability for
Clinical Laboratory Testing and In Vitro Diagnostic Test Systems,
Ellyn S. Beary, Editor, May 2001, U. S. Department Of Commerce,
National Institute of Standards and Technology.
9. Powers, Donald M.,
Regulations and Standards - Traceability of assay calibrators: The
EU's IVD Directive raises the bar, IVD Technology, July-August,
2000.
|