|
Esther
Reichstein, Ph.D.
Director, Immunochemistry and Technical Support, DPC Instrument
Systems
Division, Flanders, NJ, USA
Download as a
PDF here
The quality of patient care depends on the quality of all the
information that a physician uses in making treatment decisions. As
clinicians come to rely more and more on biochemical markers for
early detection, diagnosis, and prognosis, the accuracy of the
clinical laboratory result is an increasingly important
component of quality patient care.
In the laboratory, the interval between the time a patient�s
blood is drawn until the result is reported to the physician can be
divided into three phases: preanalytical, analytical, and
postanalytical.
The preanalytical phase involves collecting the specimen and
preparing it for analysis. This includes everything from the time
the order is received until the sample is ready for analysis (Table
1).
The analytical phase includes all the steps involved in the
actual analysis of the sample. The postanalytical phase consists of
reviewing the results for anomalies, reporting the results to the
physician, and storing the sample.
Table 1. Components of the preanalytical
phase.
Historically, clinical laboratories have focused on the
analytical phase when monitoring the quality of results. This has
involved monitoring assay performance, lot-to-lot variation and
other characteristics, typically through the use of statistical
quality control. However, it is estimated that 60 percent of errors
in the laboratory occur in the preanalytical phase1, so it is
worthwhile to consider common sources of these errors.
Sources of
preanalytical errors
Several factors related to the status of the patient can affect
the quality of the analytical result. In certain disease states,
substances in the patient�s blood may interfere with the analytical
process; a common example of such an interfering substance is
elevated bilirubin. In addition, the patient may be taking
medications that can cause assay interference. The patient may also
have circulating endogenous antibodies (heterophilic antibodies,
autoimmune antibodies and others) that can interfere with
immunoassays. The laboratory cannot control these sources of error
but it must be aware of them and consider them whenever the
analytical result seems inappropriate.
Most preanalytical errors are, in fact, associated with the
quality of the sample. Haemolysis is the most common cause of
error. Other causes include insufficient sample volume, inadequate
or incorrect labeling of the tube, and poor clotting of serum
specimens. Since many of these sources of error are the result of
specimen mishandling, attention to the proper procedures for
handling blood collection tubes is critical to result quality. Each
step in the process of preparing the sample for analysis is crucial
for maintaining sample integrity. Lack of adherence to proper
technique can result in haemolysis, or in incomplete clotting
leading to interference from fibrin. It is very important that
laboratories carefully follow the recommended procedures supplied
by the manufacturer of the collection tubes.
Fibrin
interference
Residual fibrin, long recognized as a possible interferent in
the clinical laboratory, may be present as a result of improper
specimen handling during and after collection. It can be present in
primary tube samples either as a visible clot, which may physically
occlude the instrument sample probe or, more insidiously, as an
invisible microfiber or as strands. Fibrin strands, though
invisible, may directly affect some assays, especially immunoassays
2-5. Unlike interference from heterophilic antibodies or rheumatoid
factor, fibrin interference is usually not reproducible and
disappears with time as the fibrin settles out of the sample.
Care taken during the preanalytical phase can help to reduce the
presence of fibrin strands in the processed specimen. Important
considerations in the preanalytical phase that can have an effect
on fibrin formation are shown in Table 2.
Table 2. Preanalytical phase
considerations that can affect fibrin strand formation.
-
Recognition of disease or therapy
that may affect clotting time
-
Selection of the appropriate tube
type
-
Collection sequence when multiple
tubes are collected
-
Collection tube mixing
-
Time allotted for
clotting
-
Centrifugation
-
Transportation to the
instrument
Recognition of
factors affecting coagulation
The preanalytical phase begins with the patient and the
recognition that diseases or therapies may impact on specimen
clotting. A number of conditions and treatments can affect how
readily clots form. Pregnant women and dialysis patients, for
example, often exhibit prolonged clotting times, and many patients
treated for a variety of conditions receive anticoagulants which
inhibit clotting. It is important for laboratory personnel to be
aware of these possibilities and, whenever possible, to obtain
information about the patient. If the ability of the patient�s
blood to form clots is inhibited, specimen collection must be
handled accordingly: either a plasma specimen can be used so that
clot formation is not necessary, or increased time can be allowed
for clot formation during specimen processing. If these steps are
not taken, fibrin strands may form after specimen processing and
compromise result quality.
Sample collection
The specimen collection protocol and attention to technique can
have a major effect on the quality of the analysis. Problems often
occur in critical care situations, where specimens may be collected
by procedures other than the usual venepuncture. Using an
intravenous (IV) catheter or a syringe for blood collection
followed by transfer of the blood into a collection tube can result
in a higher incidence of haemolyzed samples than is commonly seen
using routine venepuncture 6,7. These procedures can also affect
the rate of clot formation. IV lines typically contain heparin to
prevent clots from occluding the line, and the heparin can affect
the ability of the sample to clot when it is transferred to a serum
tube. Even if plasma is used for testing, sample collection by
syringe may allow microclots to begin forming before the blood is
transferred to a tube containing anticoagulant.
In the case of multiple-tube collection during a single
phlebotomy, the order of collection as described in the collection
tube product labeling should be observed; the device for puncturing
the tube stopper can become contaminated with additive from the
previous tube. Serum tubes without additives should be drawn first
followed by those with clot activator. Plasma tubes should be drawn
after serum tubes since small amounts of heparin or EDTA
contaminating a serum tube may slow clotting.
It is important to mix tubes thoroughly immediately after
collection. Most tubes contain additives that must be completely
dispersed throughout the tube to be effective. Blood drawn into
tubes containing heparin, for example, may begin to form microclots
because the localized anticoagulant concentration may be low in
some regions of the tube prior to mixing. Adequate mixing is
important even in serum tubes. Current serum tubes are plastic
rather than glass and contain an additive to enhance clotting. If
these tubes are not mixed well, clot formation will not be
adequately initiated and will take much longer to complete.
Sample processing
Serum tubes may not be allowed adequate clotting time before
centrifugation. BD (Becton, Dickinson and Company) recommends a
clotting time of 60 minutes for serum tubes and 30 minutes for gel
barrier tubes, for example. Under pressure to provide a rapid
result, laboratories all too often shorten the time allotted for
the specimen to clot. This increases the likelihood of fibrin
strand formation after centrifugation and separation, particularly
if the sample is slow to clot because the patient is pregnant or on
anticoagulant therapy.
Adequate centrifugation is required to ensure that all clots or
fibrin strands are removed. For primary tubes to work optimally
with sampling probes on automated instrumentation, the upper
surface of the barrier gel or clot should be horizontal. Use of a
fixed-angle centrifuge rotor and the resulting slanted clot or
barrier gel can result in fibrin strands being disturbed when the
tube is removed from the centrifuge and placed upright. In
addition, level-sense errors are more likely to occur when the
volume of serum is low in tubes spun in fixed-angle rotors.
Finally, care must be taken in the transportation of the sample
to the analyzer. If samples are transported from a remote site,
centrifugation in the laboratory prior to analysis is a good
practice. During transportation over long distances, there are many
opportunities for the sample to be disturbed and for fibrin strands
to be resuspended. Consequently, centrifugation in the laboratory
can prevent analytical problems, even if the specimen was
originally centrifuged at the collection site. Even with samples
that have been transferred to a secondary transport container,
centrifugation before analysis can often reduce problems from
fibrin strands that may have formed after the initial processing.
An alternative to centrifugation is the use of separation filters
that may be inserted into the collection tube or secondary tube to
capture fibrin strands and force them to the bottom of the
tube.
Conclusion
Many of the steps of sample handling prior to analysis can
affect the quality of the sample, and deviation from recommended
best practices can lead to erroneous results. While fibrin can
never be totally eliminated from primary tubes, care taken during
the preanalytical phase can minimize errors. Such a precaution will
reduce the necessity of repeating tests or redrawing the patient,
and contribute to overall cost savings and improved patient
care.
References
- Bonini P, Plebani M, Ceriotti F, Rubboli F. Errors in
laboratory medicine. Clin. Chem. 2002; 48:691-8.
- Nosanchuk JS. False increases in troponin I attributable to
incomplete separation of serum. Clin. Chem 1999; 45:714.
- Beyne P, Vigier JP, Bourgoin P, Vidaud M. Comparison of single
and repeated centrifugation of blood specimens collected in BD
evacuated blood collection tubes containing a clot activator for
cardiac troponin I assay on the ACCESS analyzer. Clin. Chem. 2000;
46:1869-70
- Ooi DS, House AA. Cardiac troponin T in hemodialyzed patients.
Clin. Chem. 1998; 44:1410-6.
- Roberts WL, Calcote CB, De BK, Holmstrom V, Narlock C, Apple
FS. Prevention of analytical false-positive increases of cardiac
troponin I on the Stratus II analyzer. Clin Chem. 1997;
43:860-1
- Kennedy C, Angermuller S, King R, Noviello S, Walker J, Warden
J, Vang S. A comparison of hemolysis rates using intravenous
catheters versus venipuncture tubes for obtaining blood samples. J.
Emerg. Nurs. 1996; 22:566-9.
- Bush V, Green S. Managing Pre-analytical Variability in
Chemistry. ASCP Fall 2001 Teleconference Series; Program No: 9080,
November 14, 2001
|