|
Bella Ho,
Hong Kong Accreditation Service
Download as a
PDF here
Introduction
ISO 15189:2003 is an international standard developed
particularly for the medical laboratories. Though it is based upon
ISO/IEC 17025:1999 and ISO 9000:2000, it is a standalone standard
for medical laboratories with a title particularly referred to
"quality and competence". The ISO 15189 requirements are however,
harmonised with those of ISO/IEC 17025. Under the International
Laboratory Accreditation Cooperation (ILAC) Multilateral Mutual
Recognition Arrangement (MLA), accreditation of medical
laboratories against ISO 15189 and ISO/IEC 17025 are both
acceptable.
The standard, since its publication in 2003, is gaining more and
more acceptance by accreditation bodies worldwide as the standard
for medical laboratories and has been adopted as the accreditation
criteria used by many economies, including New Zealand, Canada,
Israel, Hong Kong, Thailand, etc. It is also known that many
economies including Malaysia, China, Japan, are also planning to
start accreditation of medical laboratories using this new
standard. A number of accreditation bodies currently offering
accreditation for medical laboratories against ISO/IEC 17025 are
also planning to start using this new standard in the next few
years.
The Hong Kong Accreditation Service is among one of the
accreditation bodies first adopting the use of ISO 15189 to
accredit medical laboratories. As the accreditation of medical
laboratories is a new area of service offered by the Hong Kong
Accreditation Service, application documents have been prepared to
guide the laboratories in seeking for accreditation. Though ISO
15189 has been developed based on ISO/IEC 17025 and that the
essential requirements in ISO/IEC 17025 are also included in ISO
15189, there are differences in emphasis and the latter has been
written in a language more familiar to the medical community.
For those economies where the medical laboratories are not yet
accredited to any ISO standards, ISO 15189 represents a good start.
For those that have been using ISO/IEC 17025 for accrediting
medical laboratories for some years, accreditation to ISO 15189 is
a lot easier and the key is to know the difference. A detailed
comparison of ISO/IEC 17025:1999 and ISO 15189:2003 is given in
Annex A.
Management
Requirements
Basically the management requirements as stipulated in ISO 15189
and ISO/IEC 17025 are similar except that a new clause of
"Continual improvement" has been introduced in ISO 15189. The minor
differences found in the wordings of the management requirements of
ISO 15189 and ISO/IEC 17025 are in fact further elaboration on the
actual application of the standards requirements and the
expectation of quality service. The essence is the same in the two
standards.
"Continual improvement" is a requirement from ISO 9000:2000
which came out one year after ISO/IEC 17025:1999. The aim of
continual improvement of a quality management system is to improve
the service provided and to enhance the satisfaction of service
users.
Opportunities for improvement could be identified from various
sources such as feedback from service users, audits, management
review, etc. Despite having a separate clause in ISO 15189, the
concept of continual improvement is embedded in the quality
management system required under ISO/IEC 17025. Notwithstanding
this,Clause 4.12.4underContinual improvementrequires laboratory
management to implement quality indicators such as turnaround time,
blood usage, etc. for systematically monitoring and evaluating the
laboratory's contribution to patient care, this is a requirement
specific for medical laboratories.
Technical
Requirements
While there is no major difference in the basic principles
behind the technical requirements of the two standards, ISO 15189,
in a way, has additional requirements that are specific to the
medical testing laboratories and emphasis has been put on patient
care. The following focussed on the differences of requirements
between ISO 15189 and ISO/IEC 17025 and the practical application
of the new standard.
Personnel
Requirement
A much-expanded section has been devoted to describe the
responsibilities and functions of the laboratory director(s).
Clause 5.1.3 stated that "the laboratory shall be directed bya
person or personshaving executive responsibility and the competence
to assume responsibility for the services provided". This implied
the acceptance of team management with responsibilities and
competence emphasised.
The standard does not specify whether the laboratory director
should be medically qualified or the academic or professional
qualifications one must attain for fulfilling the role. The
academic or professional qualifications required of a laboratory
director or directors would need to be considered according to
local situation where local application documents would be helpful.
This is particularly valid for some economies where only limited
number of pathologists or highly academically qualified personnel,
are available. Factors to take into consideration include, but are
not limited to, the set up of the whole medical system, the local
education system and opportunities, public expectation, local
economy and whether the accreditation scheme is voluntary or
mandatory. While balancing all the factors, one important feature
that an accreditation scheme has to maintain is the quality service
that an accredited laboratory could offer. The "competence" of the
personnel and the management's commitment are the key contributing
factors to quality service, which should not be compromised due to
availability of pathologists.
This also leads to the consideration of whether visiting
pathologists who may have no responsibilities for the daily
operation of the laboratory could provide adequate medical
coverage. There is no fast and hard rule on what is adequate.
Accreditation bodies would need to be flexible when assessing a
laboratory's competency while placing priority on patient care. It
should also be recognised that for the interpretation of certain
tests, the results of examinations have to be evaluated in
conformity with the clinical information available regarding the
patient and for these tests, the input from pathologists is
inevitable.
Continual
Professional Development
Though ISO/IEC 17025 also requires the laboratory to identify
training needs of personnel and to provide relevant training, the
emphasis on continual education is much greater in ISO 15189. The
requirement of continual professional development for all staff,
including those making professional judgements, is mentioned more
than once in ISO 15189. There is also particular requirement for
staff to receive relevant training in quality assurance and quality
management as well as health and safety.
Continual professional development is in fact not uncommon to
the medical community, who always realise the need to keep up with
advances in science. Seminars, workshops, scientific conferences,
journal clubs, case studies, etc. are often organised by scientific
or professional organisations; laboratory personnel would not have
difficulties to fulfil this requirement except documentation is now
required. Support and encouragement from the management would be
required for active participation in such activities. To provide
continuing education program to staff at all levels is a
responsibility of the laboratory management.
Safety
Requirement
There is no particular safety requirement mentioned in ISO/IEC
17025 though assessors may still attend to safety issues during
assessment, it may not always be possible to raise non-conformities
against ISO/IEC 17025 on unsafe conditions observed.
Laboratory safety is important for the medical testing
laboratories, not only to the laboratory personnel, but also to the
patients, visitors and the general public e.g. in case of SARS
episodes in 2004 which have been related to laboratory exposure.
Hence ISO 15189 not only requires laboratory personnel to be
trained in safety issues, the new standard also requires patients,
employees as well as visitors, including those engineers for
repairing work, to be protected from potential risk. Attention is
also drawn to the safety for the carrier and the general public
during transportation of samples. Safety requirements are covered
underClauses 5.1 Personnel, 5.2 Accommodation and environmental
conditions, 5.3 Laboratory equipment, 5.4 Pre-examination
procedures and 5.7 Post-examination procedures.
ISO 15190 is a related standard on requirements for safety in
medical laboratories, first published in October 2003. The standard
is not yet included in ISO 15189 as a normative reference because
of its late publication, but medical laboratories meeting the
requirement of ISO 15189 are expected to refer to this safety
standard.
Laboratory
Equipment
A definition for laboratory equipment has been provided in
NOTEunder Clause 5.3 Laboratory Equipmentto include instruments,
reference materials, consumables, reagents and analytical systems.
This definition broadens the calibration and maintenance programme
of equipment to cover demonstration of proper function of reagents
and analytical systems.
Preventive maintenance is in fact a term more appropriate than
calibration to most medical equipment and analytical systems. Much
of such equipment is maintained or calibrated, if required, by
manufacturers which are not accredited calibration laboratories. It
would be impractical at this stage, to require all maintenance and
calibrations to be conducted by accredited laboratories.
Nevertheless, proper maintenance and calibration, despite being
carried out by manufacturers, has to be insisted upon, wherever
necessary. Where calibration of standard laboratory equipment e.g.
temperature monitoring device or volume measuring equipment, is
involved and such equipment affects quality of results,
calibrations should be conducted by accredited calibration
laboratories.
Pre-examination
Procedures
Samples in non-medical testing laboratories are often tested as
received. Though ISO/IEC 17025 also has requirementson Sampling,
they are applicable only for laboratories responsible for sample
collection. The requirements now covered in5.4 Pre-examination
procedureof ISO 15189 are applicable to all medical testing
laboratories.
All medical testing laboratories are required to provide a
primary sample collection manual for service users, which include
information for patients, sample collectors, and sample reception
staff. Though sample collection may be carried out by nursing staff
in hospital wards or by medical doctors in clinics, the
responsibility of proper sample collection falls onto the
laboratory. Apart from providing the necessary instructions, the
laboratory is also expected to have frequent communication and
liaison with responsible personnel to ensure that the instructions
are understood and followed.
Measurement
Uncertainty and Traceability
Measurement uncertainty is only mentioned inClause 5.6.2in ISO
15189 where there is a conditional statement of "where relevant and
possible". It is understood that the application of Measurement
Uncertainty for medical testing is still under development. There
is much concern among the medical laboratories that there is not
enough guidance of estimating uncertainty in the medical field. A
classical ISO approach to measurement uncertainty may not be
appropriate for medical testing.
Though it is generally acknowledged that the results produced by
medical laboratories should be traceable to reference materials
and/or reference measurement procedures of higher order, whenever
these are available, it is inevitable that there is a lack of
reference materials/reference measurement procedures for medical
testing. To tackle this issue, an agreement between Bureau
International des Poids et Mesures (BIPM), International Federation
of Clinical Chemistry and Laboratory Medicine (IFCC) and
International Laboratory Accreditation Corporation (ILAC) has been
signed, establishing the Joint Committee on Traceability in
Laboratory Medicine (JCTLM).
The JCTLM has published a list of about 150 reliable,
internationally recognized Certified Reference Materials of higher
order on the websites of BIPM and IFCC. Reference Materials and
Measurement Procedures included in this category are those that
provide values that are traceable to the SI units; e.g.
electrolytes, enzymes, drugs, metabolites and substrates,
non-peptide hormones and some proteins.
Another list will soon be published, on the international
conventional reference materials, i.e. where the measurand(s)
is/are not SI-traceable and/or no internationally recognized
reference measurement procedure is available; e.g. WHO reference
materials for coagulation factors, nucleic acids, and some
proteins.
At the present stage, medical laboratories are encouraged to
consider factors contributing to uncertainty of results where
possible and relevant, particularly in the discipline of clinical
chemistry. A practical and reasonable approach to Measurement
Uncertainty would be adopted when accrediting medical
laboratories.
External Quality
Assurance Programmes
Clause 4.2.2of ISO 15189 requires that the quality management
system shall include internal quality control and participation in
organised inter-laboratory comparisons, which is a clause not found
under the management requirements of ISO/IEC 17025. An expanded
section on inter-laboratory comparisons is included underClause 5.6
Assuring quality of examination procedures.
Though a number of well-established external quality assurance
programmes (EQAPs) for medical testing are available worldwide,
there are always some tests that are not covered by any programmes,
e.g. SARS detection. Besides, not many of these programmes are
accredited and whether they are in substantial agreement with
ISO/IEC Guide 43-1 as required inClause 5.6.4of ISO 15189, are not
clear. Some of these programmes are highly reputed, yet they are
expensive and their status is not clear. If locally organised
inter-laboratory comparisons are available and could be
demonstrated to be in substantial agreement with ISO/IEC Guide
43-1, participation in such local schemes should be equally
acceptable for accreditation purpose. There is, however, general
feeling that sector specific requirements for proficiency testing
providers, particularly for medical testing, is required.
For medical testing, it is generally accepted that the frequency
of participation in EQAPs should be much more frequent than the
minimum frequency of once per four years as stated by ILAC. There
is no specific guideline laid down for medical testing. However,
the medical sector does expect a higher frequency of participation
for quality service. The test areas to be covered in the quality
assurance programmes are, however, usually considerably larger than
the non-medical testing laboratories. The participation in more
than one programme is often required for adequate coverage. Hence
accreditation bodies could take a practical and reasonable approach
to the requirement of participation in inter-laboratory
comparisons, taking into consideration the scope of service
provided by the laboratory.
Establishment of
Alert/Critical Intervals and Turnaround Time
Establishment of critical/alert levels for all examinations and
turnaround time that reflect clinical needs are specific
requirements for medical testing inClause 5.8 Reporting of results.
For non-medical testing results, the limits are often those
specified in a certain standard or legislation. Laboratories
interpret their results according to these defined limits or
client's specifications. For medical testing results, the
alert/critical intervals could be dependent upon population, sex,
age or method; hence laboratories would need to determine their own
alert/critical levels according to the clinical needs. This need
and requirement is well understood by the medical laboratories.
There is no definition provided in ISO 15189 regarding
turnaround time. It could be interpreted as time of sample
collection to receipt of examination results by the requester or
when samples are received by the laboratory to issue of results.
Arguments may arise when sample collection is not under the direct
control of the laboratory, for example, in case of sample
collection in hospital wards. Both interpretations would be
acceptable provided there is justification that the defined
turnaround time reflects clinical needs. It should also be noted
that there is a requirement to monitor the transportation of the
samples to ensure that the samples are received within an
appropriate time frame. The essence of this "turnaround time"
requirement is on clinical needs and patient care.
Conclusion
The requirements specified in ISO 15189 are not difficult to
implement. Though the essence of ISO 15189 and ISO/IEC 17025 are
the same, the way that the requirements are presented in ISO 15189
are more easily understood by the medical community and are
considered as more relevant and appropriate to laboratory medicine.
Though there are areas included in ISO 15189 which are unfamiliar
to the medical community e.g. MU and traceability, the concepts are
not totally rejected and efforts have been put in developing
further guidance to the medical community. It is envisaged that ISO
15189 would gain wide acceptance by the medical community and it
would soon be used as the prime standard for accrediting medical
laboratories.
Annex A
Comparison of management and technical requirements in ISO 15189
and ISO/IEC 17025
|