|
Yolanda B. de Rijke

Introduction
In some circumstances, a creatinine check at home
instead of a complete clinical control could be sufficient for
patients with renal disease. Home monitoring of serum creatinine
concentrations could, for instance, be useful during insecure
periods after kidney transplantation, or to keep a close watch on
possible disease recurrences. Below we describe a pediatric case of
recurrent hemolytic-uremic syndrome (HUS), in which we would have
liked to monitor creatinine levels at home in order to restrict the
number of hospital visits.
Point-of-care testing (POCT) for creatinine
concentrations offers the advantage of providing a result within
minutes outside the hospital, which potentially enables faster
diagnosis and management of HUS and other conditions threatening
kidney function. Obviously, POCT devices employed at home must be
easy to use in daily practice by the patient or his/her parents.
Therefore, it was considered essential that tests could be carried
out with a small volume of capillary blood. Notably, POCT systems
used at home should produce results comparable to reference testing
in the laboratory. To that purpose, we evaluated two different POCT
systems for creatinine measurement (Roche Reflotron Plus and Abbott
i-STAT) versus the routine test used in the laboratory of the
Erasmus MC-Sophia Children�s Hospital, in 20 children displaying an
extensive range of creatinine blood values.
Children with atypical HUS have frequently been
referred to the pediatric clinic in Rotterdam. This syndrome is
characterised by microangiopathic hemolytic anemia,
thrombocytopenia, and renal failure (1). A representative case
described below gave rise to the present data.
Case history
At the age of four months, a boy presented with
atypical HUS for the first time. Plasma exchange resulted in
temporary clinical improvement, but the HUS recurred after a few
months, and plasmapheresis could not prevent him from progression
to end-stage renal failure. Peritoneal dialysis was started, yet
after two years this was switched to hemodialysis due to a
devastating peritonitis. Bilateral nephrectomy was performed at
four years of age because of therapy-resistant hypertension.
Finding a suitable deceased kidney donor was
difficult due to the patient�s homozygosity on HLA-A, -B and -DR
alleles. At the age of five years, the boy received an adult
deceased donor kidney graft. Post-transplant, serum creatinine
values quickly normalised to circa 25 �mol/L (see Figure 13.1.).
Ten days post-transplant he showed perforation of the colon
ascendens, for which a temporary stoma was applied. One month
post-transplant, an acute rejection occurred, which was
successfully treated with methylprednisolone. Subsequently, with
monthly intervals, two episodes of bacterial gastroenteritis with
increased serum creatinine concentrations due to dehydration, and a
primo Epstein-Barr virus infection occurred. A quiet period of ten
months ensued.
At the age of 6 1/2 years the patient displayed a
sudden increase in serum creatinine concentration, along with
hematuria. A graft biopsy was suspect for recurrence of HUS. The
boy was treated by plasma exchange, daily during the first two
weeks, followed by tapering in frequency, and after three months he
was switched to weekly plasma infusions. During this treatment,
serum creatinine values suddenly deteriorated, after which
hemodialysis was restarted.

Figure 13.1. Case
history: patient serum creatinine concentrations in time. Each data
point represents a single serum creatinine concentration obtained
by routine laboratory measurement (Y-axis) at the age indicated
(X-axis). Bold horizontal lines represent the reference
range.
At seven years of age, a second transplantation
was performed using a kidney from the patient�s mother. After
problems with urinary drainage followed by reimplantation of the
graft ureter, serum creatinine concentration settled at about 120
�mol/L. However, eleven months post-transplant another sudden,
biopsy-proven recurrence of HUS occurred, with a rise in serum
creatinine to approximately 700 �mol/L. Daily plasma exchanges
resulted in reversal of the creatinine concentration to
pre-recurrence level, and the frequency was very slowly tapered to
twice every week. Two further episodes of rising serum creatinine
concentration were seen, due to gastroenteritis and
dehydration.
Ideally, the patient would have been checked by
the physician on an almost daily basis. However, as the boy lived
at a distance of about 50 km from our clinic and went to school
normally, this was not feasible for him and his parents. To save
the family anxiety and several trips to the hospital, we intended
to optimise care by checking the boy�s blood creatinine value at
home.
Patients
We randomly included 20 outpatients of the
Division of Pediatric Nephrology and dialysis patients, aged two
months to 17.8 years, with varying creatinine serum levels. The
parents were informed about the study and granted informed consent,
provided that children resisting additional finger pricking would
not be enrolled. In addition, samples were taken from six healthy
adult volunteers, in order to obtain creatinine concentrations in
the normal range. Capillary blood for regular laboratory testing
and for the two point-of-care tests was collected at the same time.
Collection of capillary blood was performed according to the NCCLS
protocol (2). Blood was collected in heparinised capillaries and
immediately used to perform the Reflotron Creatinine and i-STAT
Creatinine/Crea tests. Another 400 �L of blood was drawn into a
heparin microtainer. Subsequently, tubes were centrifuged and
plasma was stored at room temperature for up to one hour until
routine creatinine measurement.
Primary outcome measure
The primary outcome measure was the difference
between the creatinine concentration obtained by the routine
automated assay (reference test) and creatinine POCT, in �mol/L. In
addition, ease of use in daily practice was extensively
assessed.
Creatinine tests
The Reflotron and i-STAT test were performed near
the patient, whereas the routine assay was carried out in the
laboratory.
The Reflotron Plus (Roche Diagnostics, Almere,
The Netherlands) is a table model diagnostic POCT device working on
the principle of reflectance photometry with approximate dimensions
21 � 30 � 35 cm, which uses reagent strips to measure a variety of
clinical chemistry parameters. For the Reflotron creatinine test,
32 �L of whole blood was collected in a dedicated lithium heparine
capillary to determine the creatinine concentration in �mol/L using
a Reflotron Test Creatinine strip (Roche Diagnostics, Almere, The
Netherlands) on the Reflotron Plus apparatus.
The i-STAT (Abbott Point-of-Care, Hoofddorp, The
Netherlands) is a hand-held diagnostic analyser of approximate
dimensions 5 � 6 � 21 cm, which uses cartridges to perform a range
of clinical chemistry tests. In our hands, the i-STAT creatinine
test required a minimum volume of 65 �L. Therefore, 125-�L
heparinised capillaries were used to collect whole blood for the
enzymatic amperometric measurement of the creatinine concentration
(in �mol/L) employing the i-STAT device with Creatinine/Crea
cartridges (Abbott Point-of-Care, Hoofddorp, The Netherlands). Two
highly qualified laboratory technicians of the Department of
Clinical Chemistry, who were blinded to the results of the routine
test, performed all POCT measurements.
For the routine automated creatinine test, the
enzymatic colorimetric CREA plus assay (which has been calibrated
against HPLC) was performed on plasma in the Hitachi 912 analyser
(Roche Diagnostics, Penzberg, Germany).
Statistical analysis
The linear association between the results from
the near-patient creatinine tests and the routine test was analysed
employing Pearson�s correlation coefficient. The differences
between the POCT and the reference creatinine concentrations were
analysed using the methods of Bland and Altman (3) and Passing and
Bablok (4). Over the concentration range between 50 and 500 �mol/L,
creatinine POCT was considered clinically valid when values did not
differ from the results of the routine test by more than 20 �mol/L
(see below), at a biological within-subject coefficient of
variation of 4.3% (5).
Results
For all patients, parental consent was given.
Additional finger pricking was resisted by none of the patients,
and sufficient material could be obtained from all of them.
Reflotron, i-STAT and routine creatinine measurements were
available for 20 children and six healthy adult volunteers. For the
20 children enrolled, the median age was 14.8 years (range,
0.2-17.8). Ten patients (50%) were male and ten (50%) were female.
The median age for the boys and the girls was 14.6 years (range,
0.2-17.8) and 14.8 years (range, 6.8-17.7), respectively.
Method agreement
For the Reflotron and i-STAT test, the
correlation coefficient versus the routine automated Hitachi 912
creatinine test was 0.99 and 1.00, respectively.
The Reflotron test generally produced lower
creatinine values than the routine test, both overall and for the
two outliers displaying a difference > 50 �mol/L (109 versus 167
and 167 versus 246 �mol/L) in the creatinine concentration range
< 500 �mol/L (Bland-Altman plot, see Figure 13.2.A).
The overall mean difference between Reflotron and
routine creatinine measurement was -16 �mol/L (n=26, 95% CI -30 to
-3). In the concentration range of up to 500 �mol/L and for
concentrations > 500 �mol/L creatinine, the mean difference was
-11 �mol/L (n=23, 95% CI -21 to -2) and -55 �mol/L (n=3, 95% CI
-149 to 39), respectively. By contrast, the i-STAT test mainly
generated slightly higher values in the creatinine concentration
range < 500 �mol/L, with a smaller distribution as compared to
the Reflotron test and no outliers showing differences > 50
�mol/L between i-STAT and Hitachi 912 creatinine values. However,
above 500 �mol/L creatinine the i-STAT produced substantially lower
values than the routine test (Bland-Altman plot, see Figure
13.2.B). The overall mean difference between i-STAT and routine
measurement was 4 �mol/L (n=26, 95% CI -5 to 13). In the
concentration range of up to 500 �mol/L and for concentrations >
500 �mol/L creatinine, the mean difference was 11 �mol/L (n=23, 95%
CI 6 to 15) and -52 �mol/L (n=3, 95% CI -68 to -35),
respectively.
With regard to the Reflotron test, 16 out of 26
results (62%) fell within � 20 �mol/L of the routine test results.
In the creatinine concentration range above 500 �mol/L, three
values were outside this window: differences between Reflotron and
routine measurement were -138, -54 and 28 �mol/L. In the creatinine
concentration range below 500 �mol/L, the maximum difference was
surpassed for 7 out of 23 (30%) pairs of data, with deviations
between -79 and 41 �mol/L.
With regard to the i-STAT test, 20 out of 26
(77%) of the results fell within � 20 �mol/L of the routine test
results. In parallel to the Reflotron data, three values were also
outside this window in the creatinine concentration range above 500
�mol/L: differences between i-STAT and routine measurement were
-61, -59 and -35 �mol/L. In the concentration range below 500
�mol/L, the maximum difference was surpassed for 3 out of 23 (13%)
pairs of data, with deviations of 21, 24 and 40 �mol/L.
When Reflotron and Hitachi 912 test were compared
according to Passing and Bablok (4), the slope was 0.95 (95% CI
0.87 to 1.06), and the intercept was -2.8 (95% CI -13.5 to 7.4; y =
0.95x - 2.8). When the i-STAT and Hitachi 912 test were compared,
the slope was 0.96 (95% CI 0.90 to 1.00), and the intercept was
16.3 (95% CI 9.2 to 24.6; y = 0.96x + 16.3), indicating a constant
bias. P-values from cusum tests for linearity were >
0.1.

Figure 13.2.
Bland-Altman comparison plots of creatinine concentrations
measured by (A) the Roche Reflotron versus the Hitachi 912 routine
automated test, and (B) the Abbott i-STAT versus the Hitachi 912
test. A dotted horizontal line represents the mean difference
between the values obtained by either POCT system and the routine
test, and the bold solid horizontal lines represent a -20 �mol/L
and 20 �mol/L difference.
Ease of use
Next to the difference between the creatinine
concentration measured by creatinine POCT and the routine test, we
extensively assessed the ease of use in daily practice for both
POCT systems. Although the Reflotron test required slightly less
material than the i-STAT test (32 versus 65 �L whole blood,
respectively), it turned out that measurement could be carried out
using only a drop of blood and yielding a result within 2-3 min
employing either apparatus. Obviously, we consider it essential
that the intended users receive the appropriate instructions from
laboratory personnel, before any POCT device can be adequately
applied at home. Although the i-STAT cartridge system may require a
little more training, we feel that Reflotron and i-STAT could both
perform well in this situation, provided that proper instruction is
given. However, the i-STAT system will be easier to use at home: as
it seems to have been developed for use in smaller laboratories, it
can be applied as a hand-held instrument. By contrast, the
Reflotron is a table model apparatus; moreover, even in the hands
of our experienced laboratory workers, air bubbles present in the
blood capillary tended to be more problematic (by yielding markedly
decreased creatinine values) using the Reflotron device.
Discussion
In the present study, two creatinine POCT devices
were compared to a routinely used laboratory method for clinical
use in children at risk for sudden decrease of renal function, such
as those with recurrent disease, as in non-infectious HUS, or
instable kidney transplant patients. In healthy children, serum
creatinine concentrations are between 18 and 88 �mol/L (18-35
�mol/L between one and four years of age, 31-68 �mol/L between four
and 13 years of age, and 37-88 �mol/L between 13 and 17 years of
age). At a creatinine value of 500-600 �mol/L, end-stage renal
failure will have been reached in all children. To detect
significant changes in creatinine blood concentrations, a home
monitoring device should be able to detect an increase in
concentration exceeding 20 �mol/L in the range between 50 and 500
�mol/L creatinine.
Both the Reflotron and the i-STAT test correlated
well with the routine automated test, especially in the creatinine
concentration range up to 500 �mol/L. The Bland-Altman plot for the
Reflotron test demonstrates good correlation with the routine test
up to a concentration of 500 �mol/L creatinine, displaying only two
outliers with deviations > 50 �mol/L. However, above a
concentration of 500 �mol/L, it shows a considerable increase in
distribution with deviations of -138, -54 and 28 �mol/L. The
Bland-Altman plot for the i-STAT test displays a smaller
distribution up to a creatinine concentration of 500 �mol/L.
However, the i-STAT also shows an increasing distribution above 500
�mol/L, with deviations of -61, -59 and -35 �mol/L.
Both near-patient tests thus show less precise
measurement of creatinine values in the highest concentration range
(> 500 �mol/L). In the abovementioned clinically relevant
creatinine range between 50 and 500 �mol/L, however, the maximum
difference was surpassed in only 30% and 13% of cases for the
Reflotron and i-STAT test, respectively.
The Reflotron and i-STAT POCT creatinine tests
can both be performed using only a drop of blood for measurement
within 2-3 min employing either. Although we did not explicitly
evaluate the use of both POCT systems outside our hospital, the use
of the i-STAT test at home seems more feasible because of its
greater ease of use. The i-STAT device was evaluated to everyone�s
satisfaction by two inexperienced medical students at the
outpatient clinic. At present, it is already frequently used for
near-patient measurements in hospital settings, with the main
purpose of blood gas and glucose monitoring at the Intensive Care
Unit. To our knowledge, this is the first study in literature, in
which a creatinine point-of-care test has been evaluated for its
possible use at home. Hence, financial support has been given by
the Social Policy unit of the Dutch Kidney Foundation in order to
cover the reagents� expenses. Taking into account the general
underestimation of creatinine values and its somewhat smaller ease
of use, we feel that, although not tested in daily practice, the
Reflotron test is less suitable for creatinine testing at home and
more appropriate for a laboratory setting.
Our eventual aim is to be provided with
predictive information by measuring the patient�s blood creatinine
concentration at home. Although the distribution with both POCT
devices suggests that they would perform well in daily practice, it
can not be ruled out that they will achieve worse when employed by
less experienced users. For both systems, we therefore consider it
essential that the patient and/or his/her parents receive the
appropriate instructions from laboratory personnel before such
devices can be adequately used at home.
Regarding both the patient�s quality of life and
the expenses of POCT systems versus plain routine creatinine
testing, it should be considered that faster diagnosis and
management of serious conditions such as HUS might be attained
employing near-patient testing at home on a daily basis. The
additional cost of the device plus the reagents might thus be
outweighed by both the patient�s eventual benefit and the reduction
in extra expenses that might have arisen otherwise (e.g., the
higher cost of hospital treatment upon leaving such conditions
untreated for a longer time).
At the moment, patients who are suspected for
changes in creatinine values have to visit the hospital for
measurement of their serum creatinine concentration. A few
transplant patients have learned to draw capillary blood, collect
it in a microtainer and send it to the hospital by regular mail, as
previously described (6). However, this takes a few days between
drawing blood and test conclusion, resulting in cumbersome
communication. Direct measurement at home provides a creatinine
value within a few minutes, after which the result may directly be
communicated to the treating physician in the hospital by email.
For the patient, this may allow restriction of hospital visits and
fine-tuning of his or her treatment.
In conclusion, the use of point-of care
creatinine tests is a promising tool for the early diagnosis and
management of conditions threatening kidney function. The Roche
Reflotron Plus with Creatinine strips and the Abbott i-STAT with
Creatinine/Crea cartridges are good candidates for near-patient
creatinine testing in blood, yet the i-STAT system seems best for
physiologic monitoring at home given its greater ease of use.
Early recognition and treatment of declining
kidney function may lead to a better prognosis in renal
patients.
Recommended literature:
- Besbas N, Karpman D, Landau D, Loirat C,
Proesmans W, Remuzzi G, et al. A classification of
hemolytic uremic syndrome and thrombotic thrombocytopenic purpura
and related disorders. Kidney Int 2006;70:423-31.
- Clinical and Laboratory Standards Institute.
Procedures and devices for the collection of diagnostic capillary
blood specimens, 5th ed. Approved standard H4-A5. Wayne: NCCLS,
2004;24(21).
- Bland JM, Altman DG. Statistical methods for
assessing agreement between two methods of clinical measurement.
Lancet 1986;1:307-10.
- Passing H, Bablok W. A new biometrical procedure
for testing the equality of measurements from two different
analytical methods. Application of linear regression procedures for
method comparison studies in clinical chemistry, part I. J Clin
Chem Clin Biochem 1983;21:709-20.
- Ric�s C, Alvarez V, Cava F, Garc�a-Lario JV,
Hern�ndez A, Jim�nez CV, et al. Current databases on
biological variation: pros, cons and progress. Scand J Clin Lab
Invest 1999;59:491-500.
- Nauta J, Hop WCJ, Grose WFA, van der Heijden AJ,
Kist JE, Wolff ED. Improved renal transplant monitoring in
outpatient clinics. Transplantation 1989;47:715-7.
|